LX-80-15 10 Hz User Manual
52
LX-80-15 10 Hz Oceanographic Radar Sensor
4. Check that you are using the correct slave device ID
The default Modbus device ID is 1. The Modbus device ID can be changed by
connecting the instrument to the PC application (over RS-232 connection).
Connect the instrument to the PC using RS-232 connection. Open the Geolux
Instrument Configurator application and establish a connection between the
instrument and the PC. Then check the
Modbus ID
parameter and make sure that it
is the same as the slave device ID used in issued Modbus requests.
5. Make sure that there are no two devices on the bus with the same Device ID.
Modbus allows to have multiple devices connected on the same bus
simultaneously. Each device must have a unique slave device ID assigned, so that
the bus master can distinguish between the devices. If two or more devices are
assigned the same slave device ID, a bus conflict will happen and prohibit the
master to correctly communicate with the slave devices. To resolve this problem,
change the instrument’s slave device ID to a unique number through the Geolux
Instrument Configurator PC application.
6. Check that you are using the correct Modbus connection parameters (baud
rate, parity, stop bits)
The default Modbus connection parameters are 9600 bps, even parity, 1 stop
bit. These parameters can be changed by connecting the instrument to the PC
application (over RS-232 connection).
Connect the instrument to the PC using RS-232 connection. Open the Geolux
Instrument Configurator application and establish a connection between the
instrument and the PC. Then check and verify that all Modbus connection
parameters are correct.