LX-80-15 10 Hz User Manual
41
LX-80-15 10 Hz Oceanographic Radar Sensor
Name
Command
Response
Details
Concurrent
measurement
aC!/aCC!
ammm8<CR><LF>
8 values are ready within mmm
seconds
Start measurement
aM1!/aMC1!
ammm9<CR><LF>
9 values are ready within mmm
seconds
Concurrent
measurement
aC1!/aCC1!
ammm9<CR><LF>
9 values are ready within mmm
seconds
Send data - aM1
aD0!
a+f+f+f+f+f<CR><LF>
a+f+f+f+f+f<CRC><CR><LF>
+f – significant wave height H
1/3
+f – significant wave height H
S
+f – zero cross-up period T
ZUC
+f – crest period T
CREST
+f – peak wave period T
PEAK
Send data - aM1
aD1!
a+f+f+f+f<CR><LF>
a+f+f+f+f<CRC><CR><LF>
+f – minimum water level L
MIN
+f – maximum water level L
MAX
+f – average water level L
MEAN
+f – median water level L
MED
Aditional data
aD2! .. aD9!
a0<CR><LF>
a0<CRC><CR><LF>
No values are returned for
additional data.
Start measurement
aM2!/aMC2!
ammm3<CR><LF>
3 values are ready within mmm
seconds
Concurrent
measurement
aC2!/aCC2!
ammm3<CR><LF>
3 values are ready within mmm
seconds
Send data - aM2
aD0!
a+f+f+f<CR><LF>
a+f+f+f<CRC><CR><LF>
+f – significant wave height H
S_M0
+f – zero up-crossing period
T
ZUC_SPEC
+f – crest period T
CREST_SPEC
Additional data
aD1! .. aD9!
a0<CR><LF>
a0<CRC><CR><LF>
No values are returned for
additional data.
Continuous
measurement
aR0! .. aR9!
a<CR><LF>
a<CRC><CR><LF>
The continuous mode is not
supported
Get measurement
unit for level
aXGLUN!
a+d<CR><LF>
+d – measurement unit for level
0 – mm
1 – cm
2 – m
3 – in
4 – ft
Set measurement
unit for level
d!
a+d<CR><LF>
+d – measurement unit for level
0 – mm
1 – cm
2 – m
3 – in
4 – ft
Summary of Contents for LX-80-15