Model TA502A
The Setup Check and Setup Modes display identical information with the following
exceptions:
•
The Setup Check Mode allows the user to view the operating parameters of
the Model TA502A, whereas the Setup Mode allows the user to change
these parameters.
•
Entering the optional Password is only available in the Setup Mode.
•
The Inhibit Mode may only be entered from the Setup Mode. If the Inhibit
Mode is entered, the unit will remain in the Inhibit Mode until the Mode/Select
switch is pressed.
•
After the Setup Mode is complete, the TA502A will enter the Setup Check
Mode to view the selected parameters.
NOTE:
The Setup and Setup Check Modes cannot be entered if the unit is in alarm
or fault. If the field device is in calibration, calibration check, or test gas
mode, the Setup and Setup Check Mode cannot be entered.
During the Setup Mode, the operator will be allowed to select options. The selection
procedure is the same for most of the options. Pressing the Mode/Select switch
toggles the available choices. When the display has indicated a choice for five
consecutive seconds without the operator pressing the Mode Select switch, the
Setup routine will accept that selection and move on to the next option available.
These modes will activate the
SETUP
LED, the
CAL-OC
output, and the
CALBUSS
.
NOTE:
Before entering the Setup Mode to make changes, the user should become
familiar with the block diagram on the following page. Filling out the form will
aid the user during the selection process in the Setup Mode.
Section 4.3 is provided to aid the operator in selecting during the Setup Mode. It is
recommended that the operator fill-in the selections in the proper blanks and then
use this page as a reference while programming the Model TA502A. The option entry
indicates the order of options in the Setup Mode. To the right of the option entry is a
description of the choices that are available for that option. More information about
making each selection is provided in the pages that follow.