Model IR400 MODBUS
5
1.3 IR400 Command Register Locations
Register
Address
(Hex)
Parameter
Function
Data
Type
Data Range Acces
s
0000H
Analog Output
0-20 mA analog
output
Numeric
Value
0-65535 Dec,
(scale to range
0-21.7mA)
R
0001h
Operating Mode
Set/View operating
mode
Bit Map
See
Table
R/W
0002h
Error Status
View present error
Bit Map
See
Table
R
0003h
Gas ID
Type of gas
Numeric
Value
See
The
register can be
written with the
index numbers
0 to 7 or with
the Gas ID.
The Gas ID
codes will not
change in
future versions
and are
consistent
across General
Monitors
products. To
read the Gas
ID use register
0x008D.
Table 9
R/W
0004H
Model Number
View Model ID
Numeric
Value
2104 , etc
R
0005h
Software Rev
Software Revision ID
2 ASCII
characters
A, B, etc.
R
0006h
Gain
Shows detector gain
Numeric
Value
0-5000
R
0007h
Cal IO type
Set/View Cal IO type
code
0 – LED switch, 1
–manual
solenoid , 2 -
ARGC
R/W
0008h
Solenoid state
Turn solenoid
ON/OFF, read
solenoid state
Code
10 – ON, 20 –
OFF, 30 –
disabled
R/W
0009h
Modbus unit
address
Set/View Modbus
address
Numeric
Value
1-247 decimal
R/W
000Ah
Adjusted ratio
Adjusted ratio
Numeric
Value
0-65535
R
000Bh
Modbus Baud
Set/View Baud Rate
(2400, 4800, 9600
Code
0, 1, 2, 3
R/W
Summary of Contents for IR400
Page 2: ......
Page 15: ...Model IR400 11 Figure 5b IR400 Mounting Dimensions with a European Box ...
Page 20: ...Model IR400 16 Figure 7 Wiring Diagram from IR400 to Control Equipment ...
Page 41: ...Model IR400 37 9 6 Engineering Documentation ...
Page 45: ...Model IR400 HART Field Device Specification for Point IR Gas Detector ...
Page 78: ...Model IR400 Infrared Point Detector for Hydrocarbon Gas Applications Modbus programming guide ...