Model IR400 MODBUS
21
1.4.61 Calibration Running Time in Seconds, Hi Word (0x00D8)
This register reads the high word of the running time in seconds when the calibration
event occurred. This time is in seconds since January 1, 2000.
1.4.62 Calibration Running Time in Seconds, Low Word (0x00D9)
This register reads the low word of the running time in seconds when the calibration event
occurred. This time is in seconds since January 1, 2000.
1.4.63 Calibration Clock Time: Year, Month (0x00DA)
These registers are described in Table as item number 1.
1.4.64 Calibration Clock Time: Day, Hour (0x00DB)
These registers are described in Table as item number 2.
1.4.65 Calibration Clock Time: Minute, Seconds (0x00DC)
These registers are described in Table as item number 3.
1.4.66 Calibration Code (0x00DD)
This returns 1 for zero events and 2 for calibration events.
1.4.67 Total Calibration Event Counter (0x00DF)
This reads the total number of calibration events have been stored in the unit.