Model IR400 HART
4
7.0 Status Information
The error status, which is returned via Common Practice Command #48, is shown in Table and
corresponds to Modbus register 0x02.
Byte
Bit
Description
Class
Device Status
Bits Set
LSB
0
IR Close to Low
Error
4,7
1 Negative
Drift
Error 4,7
2
IR is Low
Error
4,7
3
IR is High
Error
4,7
4
Brown Wire Short
Error
4,7
5
Low Line Voltage
Error
4,7
6
Failed to Calibrate
Error
4,7
7
Failed to Zero
Error
4,7
MSB 0
Gas
Check
Timeout
Error 4,7
1 Active
Lamp
Fault
Error 4,7
2
Reference Lamp Fault
Error
4,7
3 Heater
Failure
Error 4,7
4
FLASH Checksum Error
Error
4,7
5
RAM Checksum Error
Error
4,7
6
Excess Negative Drift
Error
4,7
7
EEPROM Checksum Error
Error
4,7
Table 2: Error Status Information
These bits may be set at power up to indicate an instrument failure. They may also be set by a
failure detected during continuous background diagnostic testing.