ORTHORALIX 9200
Service Manual
B-11
Gendex Dental Systems 4519-190-71193-7
5.
Connect the connectors coming from the ceph arm X8- X14-
X15 to the auxiliary board PA7.
6.
Fix the ceph arm (c) to the Z-carriage by means of the four
hexagonal head screws (M8x25) and rings supplied.
7.
Remove the panel (h) from the right-hand cover (a).
8.
Replace the right-hand cover (a) of the Z-carriage.
9.
Mount the ear plug support (d) into the cephalostat by means
of the two screws (e). The ear plug support has two vertical
mounting positions: please ask to the user which position
in more convenient.
10.
Mount the nasion rest (f) into the pin and fix it with knob (g).
B.5.4.2
Procedure for Orthoralix 9200 DDE
To install the ceph arm with Cephalostat follow the procedure
below (applicable only for the CEPH versions):
1.
Remove the right-hand cover (a) of the Z-carriage by un-
screwing five screws.
2.
Prepare the ceph arm close to the Z-carriage.
During the installation of the ceph arm, do not
handle it by the secondary collimator.
3.
Remove the metallic plate (i) by removing the four fixation
screws.
4.
Feed the free cable connector (b) from the arm through
the hole provided and connect to socket X29 of the host
computer board PC1. The wire must pass in the rear side of
the metal plate where the PC1 is mounted.
(03.0)E