ORTHORALIX 9200
Service Manual
F-72
Gendex Dental Systems 4519-190-71193-7
B) Orthoralix 9200 DDE
On the display the following message will appear:
HOST INPUTS READ
XX YY WW ZZ
where:
t UIF99::88BOE;;BSFGPVSCZUF )FYDPSSFTQPOEJOHUP
the different status of the PC1 system control board as specified
as follows:
bit position:
7 6 5 4 3 2 1 0
Byte XX
BIT
CONDITION
0
optoswitch Rb: 1= obscured, 0= not obscured
1
Not used
2
Not used
3
Not used
4
Not used
5
0= over voltage clamp on; 1=ok
6
Not used
7
TUBSU9SBZGFFECBDLUPDPOWFSUFSCPBSE12
Byte YY
BIT
CONDITION
0
optoswitch Ra: 1= obscured, 0= not obscured
1
Not used
2
Not used
3
Not used
4
Not used
5
Not used
6
Not used
7
TUBSU9SBZGFFECBDLUPDPOWFSUFSCPBSE12
Byte WW
BIT
CONDITION
0
optoswitch Ceph cassette: 1= not inserted, 0= inserted
1
optoswitch Y: 0= reset, 1= no reset
2
Not used
3
Not used
4
Not used
5
Not used
6
Not used
7
TUBSU9SBZGFFECBDLUPDPOWFSUFSCPBSE12
Byte ZZ
BIT
CONDITION
0
Dead Man: 0= ON, 1= OFF
1
optoswitch X: 0= reset, 1= no reset
2
Not used
3
Not used
4
Not used
5
Not used
6
Not used
7
TUBSU9SBZGFFECBDLUPDPOWFSUFSCPBSE12
(04.0)E
Printed on: 20 Apr 2012, 04:59:40 pm; Printed by: UWE.ZELLER