ORTHORALIX 9200
Service Manual
F-53
Gendex Dental Systems 4519-190-71193-7
Orthoralix 9200 DDE Ceph only
For the tension adjustment of the drive belts in ceph arm of the
Orthoralix 9200 DDE, proceed as follows:
1.
for the belt (a) (from the motor to the DMU (Digital Module
Unit), unlock the nut (b) of the belt stretcher and turn the
bolt (c).
2.
for the belt (d) (moving the ceph secondary collimator), un-
lock the nut (e) of the belt stretcher and turn the bolt (f).
3.
for the belt (g), unlock the two screws (h) and push or pull
the roller (i).
(03.0)E