ORTHORALIX 9200
Service Manual
F-54
Gendex Dental Systems 4519-190-71193-7
F.6.3
Adjusting the x-ray collimator
(A) Orthoralix 9200/Plus/Ceph
The X-ray collimator is moveable, with four selectable positions:
t QBOPSBNJD
t USBOTDBO
t MBUFSPMBUFSBM
t BOUFSPQPTUFSJPSQPTUFSPBOUFSJPS
To adjust the X-ray collimator, proceed as follow:
1.
Adjustment of the panoramic and transcan lead plates.
The X-ray field in panoramic and transcan mode shall be
IPSJ[POUBMMZBEKVTUFETPUIBUUIFZMJFTZNNFUSJDBMMZSFTQFDU
to the centre of the secondary slit: to do this, unlock the two
TDSFXT BBOEBEKVTUUIFIPSJ[POUBMQPTJUJPOPGUIFMFBET
support plate (e) (1 mm of shift of the plate is corresponding
to 3 mm on the secondary slit).
If it is necessary to change the vertical position of the X-ray
fields, unlock the two screws (b) and change the vertical
position of the lead plate (c) to adjust the vertical position
of the panoramic X-ray field; unlock the two screws (d) and
change the vertical position of the lead plate (f) to adjust the
vertical position of the transcan X-ray field.
2.
Adjustment of the ceph lead plates.
The X-ray fields in ceph mode (for the different formats
BWBJMBCMFTIBMMCFIPSJ[POUBMMZBEKVTUFETPUIBUUIFZMJFBU
a distance of about 7 mm from the film edges : to do this,
VOMPDLUIFUXPTDSFXT HBOEBEKVTUUIFIPSJ[POUBMQPTJUJPO
of the leads support plate (h) (1 mm of shift of the plate is
corresponding to 15 mm on the ceph cassette).
If it is necessary to change the vertical position of the X-ray
fields, unlock the two screws (i) and change the vertical posi-
tion of each lead plate (l) independently, so that they lie at a
distance of about 7 mm from the film edges.