ORTHORALIX 9200
Service Manual
B-24
(06.0)E
Gendex Dental Systems 4519-190-71193-7
B.6.4
Field upgrading to the “plus” version
For field upgrading to the PLUS version, follow the instructions
below:
1.
Switch off the system.
2.
Remove the panel with the mirror (a) by unscrewing the
screws (b) and disconnect the connector X31.
3.
Install the two integrated circuits (one enabling the DMF
projections and one enabling the Transcan projections) sup-
plied with the kit on the PC1 control board into the socket
D33 and D37 indifferently.
B.7
Options
B.7.1
External lights connection
A kit code 4519 105 20691 is available for external light connection.
The kit is providing control signals for two lights that must be
mounted externally to the room: one light is used to show the
system ready condition (stand-by condition) and the second one
is turned on during the X-ray exposure time.
Follow the instructions below for the mounting of the option:
1.
Remove the front panel with the mirror (a) by unscrewing
the two screws (b), and disconnect the connector X31.
(03.0)E