DIAGRAMS OF THE MONOPOLAR CURRENTS
Values measured within 3 seconds, by excluding the transients lower than 1 second, as established in IEC
60601
‐
2
‐
2. With regard to the powers, we stress the following:
The allowed tolerance is 20% for the powers higher than the 10% of the maximum power of the
current.
The measurements have to consider the tolerance of the devices used to perform the measurement.
The maximum deliverable power is 400W and because of this reason the PURE CUT current is
calibrated to not go over this limit.