The current is not suitable for the laparoscopic surgery since the automatic start/stop is not safe in this
case.
SEALING
‐
Current, to coagulate/seal vessels with diameter up to 8 mm, with activation by foot
‐
switch and
auto
matic “impedance sensing” stop. It is suitable, in open or laparoscopic surgery, to use clamps with jaws from
3 to 10 mm. The current is delivered by pushing the pedal until it stops, when tissues are coagulated/sealed,
automatically The current is suitable for the following:
To use clamps with strong mechanical pressure (starting from the 5/7 effect if the clamps have jaws of
5/6 mm.).
To use clamps with strong mechanical pressure (starting from the 9 effect if the clamps have jaws of 10
mm).
AUTO SEALING
‐
Current identical to the SEALING, with automatic “impedance sensing” start/stop.
The delivery starts, with a delay adjustable form 0,1 to 5 seconds in the GENERAL MENU (Section 2.6), when the
ends of the instruments come in contact with damp or bleeding tissues. The delivery stops either when the
surgeon opens the ends of the instrument or when tissues are coagulated. The current is suitable for the
following:
The current is suitable, in open surgery, for the same uses of the SEALING current.
The current is not suitable for the laparoscopic surgery since the automatic start/stop is not safe in this
DELIVERY OF THE BIPOLAR CURRENTS
Bipolar currents for cut with foot
‐
switched activation.
Push the yellow pedal.
The delivery is signalled by an acoustic signal (low, adjustable in the GENERAL MENU) and by the big image on
the touch screen.
Bipolar currents for coagulation with foot
‐
switched activation.
Push the blue pedal.
The delivery is signalled by an acoustic signal (acute, adjustable in the GENERAL MENU) and by the big image
on the touch screen.
Bipolar current “SEALING” for vessel sealing.
Clamp tissues between the jaws of the instrument and push the blue pedal of the foot
‐
switch until the
delivery stops automatically either when tissues are sealed or if the surgeon opens the jaws.
The delivery is signalled by an acoustic signal (acute, adjustable in the GENERAL MENU) and by the big image
on the touch screen. The sound ends with a different tone when the delivery stops.
Bipolar current “AUTO SOFT MICRO BICOAG”.
Clamp tissues between the tips of the instrument. The delivery starts/stops as under detailed.
The delivery starts, with a delay adjustable form 0,1 to 5 seconds in the GENERAL MENU (Section 2.6), when the
ends of the instruments come in contact with damp or bleeding tissues. The delivery stops either when the
surgeon opens the ends of the instrument or when tissues are coagulated. The delivery is signalled by an
acoustic signal (acute, adjustable in the GENERAL MENU) and by the big image on the touch screen.
Bipolar current “AUTO SEALING” for vessel sealing.
Clamp tissues between the jaws of the instrument.
The delivery starts and stops as above detailed for the AUTO SOFT MICRO BICOAG current. The delivery is
signalled by an acoustic signal (acute, adjustable in the GENERAL MENU) and by the big image on the touch
screen. The sound ends with a different tone when the delivery stops.
BE CAREFUL!
To use the bipolar scissors: Use only either the BLEND BICUT or the MICRO current and cut tissues
mechanically.
To reduce the, well known, phenomenon of the sticking of the tissue on the tips of instruments: See the
par: STICKING OF THE TISSUE ON THE TIPS OF THE BIPOLAR INSTRUMENTS FOR
COAGULATION.
To check the functioning of the bipolar instruments when starting the operation: See the par: CHECK OF
THE BIPOLAR INSTRUMENTS BEFORE THE USE.
THE ALARM SIGNALS DURING USE