Entering further basic program data
28
Switch functions of the limit value outputs
LV1 and LV2
Switch function 0, duration
If the limit value LV1 or LV2 has been exceeded, the output relay LV1
or LV2 switches. If the measured value falls below the limit value
LV1 or LV2 without locking, the relevant relay drops out again.
Switch function 1, impulse
If the measured value exceeds the limit value LV1 or LV2, the rele-
vant output switches for a settable time (t).
The respective output always remains switched for the set time, irre-
spective of how long the limit value has been exceeded. A new im-
pulse is only possible once the value has fallen below the limit value!
Switch function 2, interval
If the limit value has been exceeded, the respective output switches
at intervals with the settable time (t) = impulse or interval as long as
the limit value is exceeded.
Function IN1
This input is used to suppress change-over of the parameters or the
measuring points. The measurement only occurs at parameter 1 or at
measuring point 1 with parameter 1.
See section “Description of the signal inputs/outputs".
In
, select
=>
=>
Select a normally closed or normally open contact
Press "ENTER" to confirm the entry
Water meter
It is necessary to connect a water meter to
input IN2
for quantity-
dependent analysis triggering. Program the corresponding water
meter rating.
In
, select
=>
=>
Select the water meter rating
Press "ENTER" to confirm the entry
▼▲
▼▲
▼▲
Switch functions dia-
gram