Installation
10
Installation
Risks resulting from incorrect installation!
Install Testomat 2000
®
DUO at a location where it is protected
against dripping or splash water, dust and aggressive substances
– e. g. in a switch cabinet or on a suitable wall.
Information for trouble-free operation
Install Testomat 2000
®
DUO vertically and without mechanical
stress.
Install Testomat 2000
®
DUO
at a vibration-free site.
Operating Testomat 2000
®
DUO in the pres-
sure range 0.3 to 1 bar
Prior to installation, please check whether lower operating pressure is
required. The instrument is factory set for the operating range 1 to 8
bar. Remove the flow controller valve body
to operate the instru-
ment in the operating range 0.3 to 1 bar (e.g. when using an aerator
type R). This involves removing the retaining pin
from the control-
ler / filter receiver
. Subsequently use the metal bracket to remove
the controller plug
from the borehole. Subsequently remove the
flow controller valve body
and reinsert the controller plug and the
retaining pin.
Installing Testomat 2000
®
DUO
Select an installation site where the water inlet hose can be kept as
short as possible (max. 5 m).
Please leave sufficient space on the left-hand side of the instru-
ment to open the door.
Drill the mounting holes as shown in the drawing on the left.
Use three screws to attach the instrument at a suitable position in
the switch cabinet or on a wall.
NOTE
Remove for
pressure
range 0.3 to 1
bar
WARNING
!