Information menu "i"
39
Information menu "i"
In the information menu, it is possible to request active settings and
statuses of the instrument, the error history, the date for the next
maintenance and the customer service address.
Call (1)
Use the key to open the information menu "i".
Request options: Customer service, operating values, program val-
ues, error history, maintenance
Please refer to the section "Password protection and basic program-
ming" for further programming and setting details for the individual
menu items.
Operating values (3)
Display of the current values.
Program values (4)
Use the arrow keys to open the
menu item "Program values".
Press the "ENTER" key to open
the list of the set values. Press
the "ENTER" key to request the
current setting of a parameter
An asterisk indicates the se-
lected functions. (There are no
active lines.)
Error history (5)
Use the "i" and "ENTER" keys
to open the error history. The
error history is a list of errors or
statuses which have occurred
during operation. The list is
deleted after a power failure
and a new list is started.
If no errors have occurred
since start-up, the last switch-
on time of the unit is displayed,
e.g.:
POWER FAILURE
from 16.06.09 06:56
to 16.06.09 07:09
Maintenance (6)
Display of the next mainte-
nance date and the pro-
grammed maintenance interval.
It is possible to set the mainte-
nance interval in the basic
program (password protected).
Refer to the section entitled
"Maintenance" for further de-
tails on maintenance intervals.
Customer service (2)
Display of the customer service
address or e.g. a service hot-
line number.
You can freely program these
three lines in the basic progam
(password protected).