51
Check List Testomat 2000
®
Dear customers and service technicians,
This check list cannot replace your expertise or extensive experience in fault resolution. It is intended to support fast and systematic error diagnosis
and error documentation. This list does not claim to be complete. We are therefore always grateful for any advice and information you may be able
to provide. General user instructions can be found on the rear of this check list.
The Instrument Manufacturer
Block 1 / Plant and instrument data
Testomat 2000
®
Testomat
®
ECO
Plant type
Instrument type
Instrument no.
Indicator type
Software status
Pump no.
Block 2 / Error message and error history
Please mark appropriately (X)
What does your instrument’s error history display?
(“i” and “Enter” key => operating instructions)
( Error history text )
Does an error message appear on the display?
For example, “Mf. analysis”, “Low water level”, etc.
(See operating instructions: “Error messages / Troubleshoot-
ing”)
Yes
No
( Error message text )
Block 3 / Visual inspection and functional test
Please mark appropriately (X)
If applicable, values / comments
Is the instrument connected to the mains voltage specified on the rating plate?
Yes
No
Does a message appear on the display?
Yes
No
Does the instrument display a plausible measured value?
(possible manual measurement ___________ value)
Yes
No
Measured value:
Are the measuring chamber and sight-glass windows clean?
Yes
No
Are the measuring chamber and the water-carrying hoses free of leaks?
Yes
No
Is the indicator’s expiry date still valid?
(See expiry date printed on the indicator bottle)
Yes
No
Expiry date:
Has the correct indicator type been programmed?
(TH 2025 => 0.25 to 2.5 °dH = factory setting)
Yes
No
Type:
Is the water pressure within the specified range (400 ml/min)?
(See the type plate on the instrument)
Yes
No
Plant pressure:
Does the drain hose prevent the risk of backwater?
(No “syphoning effect”!!)
Yes
No
Is the drain hose free of blockages?
(Microorganisms caused by contamination, etc.)
Yes
No
Does the set flushing time/quantity of flush water ensure that only freshwater
is measured?
Yes
No
Flushing time:
Are the hoses at the dosing pump free from air bubbles?
(Operate the pump manually / Carry out a manual analysis)
Yes
No
CARRYING OUT A (MANUAL) ANALYSIS
Does the water column rise evenly up to the overflow borehole when filling the
measuring chamber (5 mm below the upper edge of the measuring chamber)?
(If not: check the water pressure, water throughput/flow controller)
Yes
No
Does the indicator pump dose correctly when starting an analysis?
(LED at the pump illuminates!)
Yes
No
No. of dosing strokes:
Have the indicator and water been mixed properly in the measuring cham-
ber after the dosing process?
Check the magnetic stirring bar! =>see maintenance manual “Adjust mode”
Yes
No
PROGRAMMING DATA / OPERATING CONDITIONS
Have the limit values been set correctly? (Within the measuring
range/according to the performance limit of the plant?)
Yes
No
Limit values:
Is the Testomat instrument always supplied with mains voltage
– except
during maintenance work/emergency situations?
(Temporary shutdown only via the “Standby” key or the “Stop” input!)
Yes
No
See the “General instructions for operat-
ing Testomat 2000
®
and Testomat
®
ECO”
Please refer to “Error messages / Troubleshooting” in the
operating instructions
for further information on error messages and possible causes
of faults.
Further functional tests (e.g. overflow detection and amplification setting => “Special function Adjust mode”) and service instructions can be found
in the
maintenance manual
.
After completing these checks, experience shows that it can be assumed that the checked functions (Block 3) are in effective working order if you
have answered all the questions with “Yes”. We recommend you to carry out these checks during each inspection or if faults occur.