Add liquids into the disposable
bag
CAUTION
Spread of hazardous and biological substances. Do not overfill
the bag. An overfilled bag might burst.
NOTICE
Adjust the disposable bag position during the filling procedure.
Folds in the disposable bag might interfere with the mixing process.
Adjustment of the disposable bag also makes sure that the maxi-
mum capacity of the bag is used.
To add liquids into the disposable bag, follow the instructions below.
Action
Step
Start filling the disposable bag through the liquid addition line (i.d. ¾
″
).
1
Stop filling when the liquid level has reached 3 to 5 cm and the bag begins
to adjust to the shape of the mixer vessel.
2
Verify that the following conditions are met:
3
•
No folds in the bag assembly
•
Minimal wrinkles in the bag assembly
Tip:
Check for wrinkles at the bag bottom.
Adjust the bag assembly inside the mixer vessel as follows:
4
•
Adjust the bottom corners of the bag assembly. If your mixer has a side
access opening, use this opening for the bag adjustment.
•
Pull the top corners of the bag assembly towards the correct position.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
231
7 Operation
7.4 Add liquid contents
Summary of Contents for Xcellerex XDM Series