Action
Step
Prepare an applicable container for collection of the drained coolant. See
Section 10.1 System specifications, on page 274 for the information on the
vessel jacket volume.
3
Connect the drain valve (1) to the collection container as applicable.
4
1
2
Open the drain valve (1).
5
Open the vent valve (2) and collect the coolant.
6
When the fluid flow stops, lift up the coolant outlet tubing (RETURN) to drain
the content of the tubing into the vessel jacket.
7
Prepare the tubing for draining the vessel jacket bottom:
8
a
Attach a Snap-tite connector to the tubing.
b
Put the end of the drain tubing into the collection container.
c
Connect the Snap-tite connector of the drain tubing to the
coolant inlet tubing (connected to SUPPLY).
See Fill the vessel jacket, on page 123 for more information about Snap-tite
connectors.
Wait until all fluid has drained out from the mixer vessel jacket.
9
128
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
5 Installation
5.4 Connect temperature control unit