![GE Transport PT878 User Manual Download Page 230](http://html.mh-extra.com/html/ge/transport-pt878/transport-pt878_user-manual_128485230.webp)
Chapter 10. Specifications
214
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
Note:
Accuracy depends on pipe size and whether measurement is one-path
or two-path. Accuracy to ±0.5% of reading may be achievable with
process calibration.
Range:
–12.2 to 12.2 m/s (–40 to 40 ft/s)
Rangeability:
400:1
Repeatability:
±0.1% to 0.3% of reading
Note:
Specifications assume a fully developed flow profile (typically 10
diameters upstream and 5 diameters downstream of straight pipe
run) and flow velocity greater than 1 ft/s (0.3 m/s).
Energy Accuracy
The accuracy of the energy measurement is a
(% of reading):
combination of the accuracy of the flow and
temperature measurement. Accuracy of 1% is
typical for calibrated systems.
Low Voltage
This unit complies with European Standard
Directive
EN61010-1 Installation Category II,
Compliance:
Pollution Degree 2.
EMC
This unit complies with EN61326:1998
Directive
CL.A Annex C. For continuous unmonitored
Compliance
operation, follow provisions of the 89/336/EMC
Compliance.
10.2 Electrical
Internal Batteries
Type:
Rechargeable batteries. (Optional NiCad 6
R
nonrechargeable alkaline available.)
Battery Life:
9-11 hours of continuous operation is typical
Battery Charger:
Input:
100-250 VAC, 50/60 Hz, 0.38 A
Memory:
1 MB datalog/site battery-backed RAM
1 MB program FLASH memory