Revision A
Transport Pro
6-5
2012659-042
Troubleshooting: Problems
Missing/Inaccurate Patient Data/Waveforms
1. Refer to operator’s manual for detailed information on patient
preparation and electrode placement.
2. Change the sensitivity level. Usually a lower percentage rectifies the
problem.
3. Test the patient cable on a known working device. If the patient cable
is defective, replace.
4. Test the patient connector with a patient simulator. If the patient
connector is defective, replace the acquisition device.
Speaker Buzz
During Alarms
1. Reboot the monitor.
2. Reseat speaker cable connection.
3. Verify that speaker and/or cable is not damaged.
4. If speaker or cables are damaged, replace the speaker FRU.
5. Unplug speaker and check speaker cable resistance.
6. If resistance is between 6 and 8 ohms, replace the main processor
FRU.
7. If resistance is outside 6 and 8 ohms, replace the speaker.
Always
1. Reboot the monitor.
2. Replace the main processor FRU.
Unresponsive Buttons
1. Reboot the monitor.
2. Connect a known good power supply.
3. Reseat the keypad harness connection.
4. Verify that the keypad harness is not damaged.
5. If the keypad harness is damaged, replace the front bezel.
6. If the keypad harness is not damaged, replace the main processor
FRU.
Summary of Contents for Transport Pro
Page 1: ...Transport Pro Patient Monitor Service Manual Software Version 2 2024579 018 Revision A ...
Page 9: ...Revision A Transport Pro 1 1 2012659 042 1 Introduction ...
Page 12: ...1 4 Transport Pro Revision A 2012659 042 Introduction Safety Message Definitions ...
Page 13: ...Revision A Transport Pro 2 1 2012659 042 2 Equipment Overview ...
Page 28: ...2 16 Transport Pro Revision A 2012659 042 Equipment Overview Theory of Operation ...
Page 29: ...Revision A Transport Pro 3 1 2012659 042 3 Installation ...
Page 33: ...Revision A Transport Pro 4 1 2012659 042 4 Configuration ...
Page 38: ...4 6 Transport Pro Revision A 2012659 042 Configuration Monitor Settings ...
Page 39: ...Revision A Transport Pro 5 1 2012659 042 5 Maintenance ...
Page 47: ...Revision A Transport Pro 6 1 2012659 042 6 Troubleshooting ...
Page 57: ...Revision A Transport Pro 7 1 2012659 042 7 Ordering Parts ...
Page 59: ...Revision A Transport Pro 7 3 2012659 042 Ordering Parts Exploded View Exploded View 622A ...
Page 62: ...7 6 Transport Pro Revision A 2012659 042 Ordering Parts Hardware ...
Page 63: ...Revision A Transport Pro 8 1 2012659 042 8 Disassembly ...
Page 83: ...Revision A Transport Pro 9 1 2012659 042 9 Checkout ...
Page 97: ...Revision A Transport Pro A 1 2012659 042 A Technical Specifications ...
Page 100: ...A 4 Transport Pro Revision A 2012659 042 Technical Specifications External Power Supply ...
Page 101: ...Revision A Transport Pro B 1 2012659 042 B Electromagnetic Compatibility ...
Page 109: ......