Revision B
SEER Light Ambulatory Recorder/Controller
3-13
2019818-008
SEER Light/SEER Light Extend Controller: Troubleshooting
Recording Start Date/Time is Not Correct
Time is Not Correct
Recording start date/time
is not correct.
Is date within
one month?
Adjust date.
See next flowchart.
Yes
No
Time is not correct.
Is it within 5 years
of use?
Call GE service.
Are the
battery electrodes
clean or properly
solder?
Clean the electrodes
or improve soldering.
Replace the PC boards.
Yes
Yes
No
No
Summary of Contents for SEER Light
Page 6: ...iv SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 7: ...Revision B SEER Light Ambulatory Recorder Controller 1 1 2019818 008 1 Introduction ...
Page 8: ...1 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 20: ...2 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 40: ...3 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 53: ...Revision A SEER Light Ambulatory Recorder Controller 4 1 2019818 008 4 SEER Light Connect ...
Page 54: ...4 2 SEER Light Ambulatory Recorder Controller Revision A 2019818 008 For your notes ...
Page 61: ...Revision B SEER Light Ambulatory Recorder Controller 5 1 2019818 008 5 Maintenance ...
Page 62: ...5 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 68: ...A 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 74: ...B 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 81: ...Revision B SEER Light Ambulatory Recorder Controller Index 1 2019818 008 Index ...
Page 82: ...Index 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 84: ...Index Index 4 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 85: ......
Page 86: ......