SEER Light/SEER Light Extend Recorder: Troubleshooting
Cannot Detect Pacemaker Pulse
Cannot detect
pacemaker pulse.
Insert the batteries while
pressing the STOP button.
Disconnect the patient cable
and touch the electrode
connector with a finger.
Does audible beep
sound?
Normal pacemaker detection circuit.
Confirm pacemaker pulse detection
by the Preview function of the
controller. Reposition the CH1 if the
pulse signals are too small.
Call GE Medical Systems
Information Technologies
Service.
No
Yes
Summary of Contents for SEER Light
Page 6: ...iv SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 7: ...Revision B SEER Light Ambulatory Recorder Controller 1 1 2019818 008 1 Introduction ...
Page 8: ...1 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 20: ...2 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 40: ...3 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 53: ...Revision A SEER Light Ambulatory Recorder Controller 4 1 2019818 008 4 SEER Light Connect ...
Page 54: ...4 2 SEER Light Ambulatory Recorder Controller Revision A 2019818 008 For your notes ...
Page 61: ...Revision B SEER Light Ambulatory Recorder Controller 5 1 2019818 008 5 Maintenance ...
Page 62: ...5 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 68: ...A 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 74: ...B 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 For your notes ...
Page 81: ...Revision B SEER Light Ambulatory Recorder Controller Index 1 2019818 008 Index ...
Page 82: ...Index 2 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 84: ...Index Index 4 SEER Light Ambulatory Recorder Controller Revision B 2019818 008 ...
Page 85: ......
Page 86: ......