January 2006
5-10
Programming the 4x40 Optica
Automatic Cleaning and
Balance Function (cont.)
3.
Manually clean the mirror as required.
Note:
Manual cleaning provides the most thorough cleaning.
4.
After manual cleaning, initiate the
Automatic Cleaning &
Balancing
cycle. (This can be done by powering the analyzer off
then on again, if the front panel softkey is disabled).
5.
To program a specific time of day at which to initiate the
Automatic Cleaning and Balancing
:
a.
Highlight either
Auto-D
or
Pacer-D
.
b.
In the
Set Time
section, input the time of day that you would
like to initiate the balance cycle in a 24-hour format (for
example, 13:30 will be 1:30 PM ).
Note:
In this mode the front panel Pacer initiation function is
disabled. Powering down and restarting the unit will have no
effect on the time programmed, however the PACER will run
on startup as is the normal function.
6.
To program an elapsed time to initialize the
Automatic Cleaning
and Balancing
:
a.
Highlight either
AUTO
or
Pacer
.
b.
In the set time section, input the elapsed time in minutes. (For
example: 720 would enable the balance cycle to run every 12
hours).
Note:
If the unit is powered down and restarted, or the Automatic
Balance is initiated from the font or using a LAN, the elapsed
time will reset.
Note:
Show Status
should read "
Enabled
".
7.
Select either
Track
or
Hold
for the analog outputs
(4-20mA/0-5VDC).
Note:
If
Track
is selected, the actual temperature of the mirror will
be transmitted. If
Hold
is selected, the last prevailing dew
point, measured before the balance cycle was initiated, will be
transmitted during the time that the balance cycle is running.
8.
Power down and restart the Optica 4x40 display analyzer for the
settings to take effect.
IMPORTANT:
For environments or a gas sample where the mirror
accumulates contamination rapidly, the use of an
inline filter is recommended. Lower flow rates will
also reduce the accumulation of contaminants
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...