January 2006
5-8
Programming the 4x40 Optica
Automatic Cleaning and
Balance Function
Note:
For Optica Analyzers with earlier versions of software, see
Appendix F
.
Optica Analyzers with version XXX software have been upgraded to
include the ability to program the PACER self-cleaning and
rebalancing cycle to run once per day at a preset time. This is referred
to as a
Clock Time Interval
. When the unit is programmed in this
manner, the front panel PACER softkey is disabled. The
Elapsed Time
Interval
initiates the PACER at a preset time after the last PACER was
run. The automatic balance cycle will always run upon power up of
the analyzers.
To program the
Automatic Cleaning & Balance Function
on the
Optica 4X40 monitor:
1.
Make sure the internal clock of the Optica has the correct time.
From the main screen:
a.
Press the
ENTER
key.
b.
Press the
DOWN
soft key until
Set Time and Date
is
displayed.
c.
Press the
>>>>
soft key.
d.
Press the
>>>>
soft key.
e.
Highlight the
Date
and
Time
fields using the tab key. Use the
keypad and softkeys to set the correct time.
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...