Installation
2-9
January 2006
Analog Outputs
Note:
When the Optica is being programmed, the analog outputs
provide 4–20mA and 0–5 VDC signals representing the
designated parameters.
•
For 4–20mA output, connect to terminals labelled
4–20 (+)
and
RTN (–)
.
Note:
The maximum load allowed for current output is 500 Ohms.
•
For 0–5 VDC output, connect to terminals labelled
0–5 (+)
and
RTN (–)
.
Note:
The maximum load allowed for voltage output is 5 mA.
Example:
Assume a temperature output, scaled to range from 0°C (Tlower) to
100°C (Tupper), with a measured actual temperature of 23°C
(Tactual):
The voltage output is calculated by:
yielding an output voltage of 1.15V.
The current output is calculated by:
yielding an output current of 7.68 mA.
Vout
Tactual Tlower
–
(
)
Tupper Tlower
–
(
)
---------------------------------------------- 5
×
=
23 0
–
(
)
100 0
–
(
)
--------------------- 5
×
1.15V
=
Iout mA
(
)
Tactual Tlower
–
(
)
Tupper Tlower
–
(
)
----------------------------------------------
20 4
–
(
)
×
4
+
=
23 0
–
(
)
100 0
–
(
)
---------------------
16
( )
×
4
+
7.68mA
=
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...