Operation
3-7
January 2006
Measuring Vapor Pressure
With
a Pressure Sensor
To measure vapor pressure with a 4-20 mA, 0-30 psia pressure sensor,
connect the equipment as shown above and make the following
entries in the Pressure Input menu section.
•
Input: Iin 4-20
•
Units: psia
•
Upper: 30.00
•
Lower: 0.00
In addition, to accommodate the pressure expansion, the following
entries are required in the Process section:
•
Status: enabled
•
Pressure: 500 (specify the process pressure with the same units
as the default pressure specified above)
Measuring Dew Point With
or Without a Pressure
Sensor
In Scenario 2, the dew point measurement requires both the sensor
pressure and the process pressure to be known. Sensor pressure can
be entered as a default value, or measured, as above, and the process
pressure must be entered in the Process Pressure field.
Example (using standard atmospheric conditions at 25°C)
P1 = 500 psia
P2 = 14.7 psia
As measured by the GEI chilled mirror dew point sensor:
Tdew2 = –40°C @ P2
e2 = 0.1283 mbar
Per Dalton’s Law of Partial Pressure:
e1 = (P2/P1) × e2 = (500/14.7) × 0.1283 = 4.36 mbar
Using standard vapor pressure equations, the actual pressurized dew
point is calculated by the Optica to be:
Tdew1 = –4.022°C
Summary of Contents for Optica
Page 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Page 11: ...Chapter 1 ...
Page 20: ...Chapter 2 ...
Page 43: ...Chapter 3 ...
Page 58: ...Chapter 4 ...
Page 80: ...Chapter 5 ...
Page 94: ...Chapter 6 ...
Page 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Page 99: ...Chapter 7 ...
Page 109: ...Appendix A ...
Page 114: ...Appendix B ...
Page 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Page 119: ...Appendix C ...
Page 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Page 122: ...Appendix D ...
Page 127: ...Appendix E ...
Page 128: ...Glossary ...
Page 130: ...Appendix F ...
Page 133: ...Appendix G ...