M
Millenium™ RO System by GE Osmonics
14693e - 31 Mar 03
4- 9
Routine Maintenance
4.2.4 Product Recirculation Check Valve Test
With Valve In RO Unit
1.
Measure TDS of Product water using a handheld meter. If RO display and TDS
measurement do not match within a small amount, replace check valve. If values
match, check valve is functioning correctly.
2.
Measure and record the Product water flow (using a flow meter or graduated
cylinder and stopwatch). After determining flow rate, slowly kink the waste line
being careful to not close off the flow completely. Measure Product water flow
again and compare against previous results. If there is a noticeable increase in
flow, the check valve needs replaced.
3.
If none of the above occurs, check valve is functioning correctly. For further
verification, proceed with the following test.
With Valve Removed From RO Unit
1.
Apply pressure in opposite direction of flow indicated on check valve. Check
pressure reading (using a gauge) and for seepage. Remove pressure from valve
and cycle it (push in direction of flow several times using care not to scratch or
damage valve). Reapply pressure in opposite direction of flow and recheck
pressure and for seepage. If pressure does not drop and no seepage is found,
valve is functioning correctly.
Summary of Contents for MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS...
Page 2: ......
Page 6: ...14693e D D 31 March 2003 This page intentionally left blank...
Page 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES...
Page 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES...
Page 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES...
Page 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES...
Page 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View...
Page 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic...
Page 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES...
Page 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES...
Page 106: ......