M
Millenium™ RO System by GE Osmonics
14693e - 31 Mar 03
5- 6
Troubleshooting
2.
Pump pressure too high.
a. Adjust pressure at pump head.
b. Check waste flow-path for obstruction.
c. See
5.1.6
d. See 5.1.7
3.
Thermal overload trips
a. Verify line voltage and motor amperage draw.
b. Disassemble pump/motor at clamp and inspect coupler for wear.
c. Pump pressure too high (see 5.2.2).
5.3
Motor Runs Hot
1.
Pump/motor binding.
a. Remove pump head and try restarting motor.
b. Inspect pump and motor coupling for wear.
c. Clean any debris or liquid from connection area.
d. See 5.5
e. Replace
pump.
2.
Fan obstructed or inoperative.
a. Verify no debris is obstructing fan vent.
b. Verify fan operation.
3.
Excessive current draw.
a. Verify that flow rates and pressures are within RO specifications.
b. Check motor/pump coupling for alignment.
c. Confirm amperage draw with meter.
d. Motor insulation shorted, replace motor.
4.
Voltage too low.
a. Check voltage against serial label and motor label, correct condition.
5.4
Water Produced While RO is Off
1.
Faulty inlet water solenoid valve.
a. Cycle valve off and on with inlet water valve switch to clear obstruction.
b. Remove valve and clear obstruction.
c. Replace
valve.
2.
Fault inlet water valve switch.
a. Verify switch is in OFF position.
b. Check switch with ohm meter, replace switch.
Summary of Contents for MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS...
Page 2: ......
Page 6: ...14693e D D 31 March 2003 This page intentionally left blank...
Page 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES...
Page 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES...
Page 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES...
Page 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES...
Page 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View...
Page 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic...
Page 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES...
Page 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES...
Page 106: ......