System Configuration
Field
Comment
Display Lead Group
Determines which group of leads is displayed.
The available values depends on which
Display Format
is selected. For example, if
3
Leads: 1x3
is selected, the available values
are:
• 3 rhythm leads
• 1st group
• 2nd group
• 3rd group
• 4th group
If either of the 12–lead display formats is
selected, this field is not available, since all 12
leads are displayed.
Page 2
Printer Leads
Identifies the default set of leads to use for
printing. The values are:
• First 6
• Second 6
• Rhythm 6
Start rhythm report on new page
Determines whether the rhythm report should
begin on a its own page.
Pace Enhancement
Increases the readability of pacemaker ECG
either by augmenting small pace pulses or by
truncating large pace pulses. If enabled, pace
enhance is done in two steps:
1.
Add a marker (1.5mV amplitude, 6 ms
duration) to the electrode signal.
2.
Limit the sum to 0.5mV in the lead signal.
Hookup Advisor
Enables/disables the
Hookup Advisor
option,
which visually indicates the quality of lead
signals. For more information, refer to
“Hookup Advisor Module” on page 61
2060026-001C
MAC™ 800
93
Summary of Contents for MAC 800
Page 8: ...8 MAC 800 2060026 001C ...
Page 26: ...Introduction 26 MAC 800 2060026 001C ...
Page 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Page 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Page 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Page 122: ...System Configuration 122 MAC 800 2060026 001C ...
Page 128: ...Maintenance 128 MAC 800 2060026 001C ...
Page 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Page 153: ......