Recording a Resting ECG
Option
Description
Reanalyze
Allows you to edit the global measurement and T-wave
dispersion. This feature is available only if:
• Audit Trail is disabled on the
Basic Setup
window
• Either measurement option (ME12 or MI12) is enabled
• Reanalysis option is selected in the
Resting ECG Setup
window
Laser Print
This feature is available only if the
Laser Printer (LPRT)
option
is enabled.
Prints the ECG report to an external USB laser printer or network
laser printer.
ADS
Toggles the anti-drift system (
ADS
) on and off.
ADS
helps reduce
baseline drift.
Main Menu
Exits the Resting ECG function and returns to the Main Menu.
Generating a Rhythm Report (Manual Recording)
The
Resting ECG
mode allows you to generate Rhythm Reports, which are printed
reports only. They do not have computer-generated interpretation or measurements,
and you cannot store them to internal memory or transmit them. Use the following
steps to generate a Rhythm Report.
1.
Prepare the patient as described in
“Preparing the Patient” on page 41
2.
Verify that the system is in
Resting ECG
mode.
If the system is not in
Resting ECG
mode, on the
Main Menu
press
F1
to select
Resting ECG
.
3.
Enter the patient data as described in
“Entering Patient Information” on page 47
.
4.
Adjust the
Speed
,
Gain
, and
Low Pass Filter
until the waveforms are configured
as desired.
For more information, refer to
“ECG Options” on page 53
.
5.
If the patient has a pacemaker, press
F5
to turn on
Pace Enhance
.
For more information, refer to
“ECG Options” on page 53
.
6.
Press
Leads
to scroll through all 12 leads.
For more information on display formats, refer to
“Resting ECG Setup” on page
7.
Press
F6
to select
More
.
8.
Press
F1
to select the appropriate
Printer Leads
option.
For more information on the
Printer Leads
option, refer to
page 53
.
9.
Press
Rhythm
to begin recording the ECG.
10.
Press
Stop
to stop the ECG recording.
If you press
Rhythm
after pressing
Stop
, the new report either begins printing
immediately on the current sheet of paper or advances to a new page, depending on
2060026-001C
MAC™ 800
57
Summary of Contents for MAC 800
Page 8: ...8 MAC 800 2060026 001C ...
Page 26: ...Introduction 26 MAC 800 2060026 001C ...
Page 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Page 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Page 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Page 122: ...System Configuration 122 MAC 800 2060026 001C ...
Page 128: ...Maintenance 128 MAC 800 2060026 001C ...
Page 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Page 153: ......