Arrhythmia Mode Recording
Function Key
Option
Description
F3
Gain
Changes the magnitude of the ECG signal on
the display or in the report. Measurement is in
millimeter per millivolt (mm/mV) and includes
the following options:
• 5 mm/mV
• 10 mm/mV
• 20 mm/mV
• 40 mm/mV
• 2.5 mm/mV
• Automatic
The larger the selected measurement, the larger
the waveform appears. Only the appearance of
the waveform changes; signal strength is not
affected.
NOTE:
If
Automatic
is selected, the system
calculates the best gain based on the
peak-to-peak amplitudes of all displayed
leads and the selected display format.
F4
Filter
Eliminates noise in the waveform by restricting
which frequencies are included. Frequencies
are measured in Hertz (Hz) and include the
following options:
• 20 Hz
• 40 Hz
• 100 Hz
• 150 Hz
Selecting a frequency eliminates signals
that exceed that frequency. The smaller the
frequency selected, the more signal is filtered
out. For example, a filter of 40 Hz displays only
signals at 40 Hz or less; signals greater than 40
Hz are ignored.
CAUTION:
INACCURATE READINGS — Using the filter
can result in a cleaner waveform, but
selecting a frequency that is too low could
alter the waveform’s morphology, resulting
in an inaccurate reading.
To avoid this, use the filter only to eliminate
excessive noise and use the highest
frequency that provides a readable
waveform.
F5
ADS
Toggles the anti-drift system (
ADS
)
on
and
off
.
ADS helps reduce baseline drift.
F6
More
Toggles through the softkey options.
2060026-001C
MAC™ 800
67
Summary of Contents for MAC 800
Page 8: ...8 MAC 800 2060026 001C ...
Page 26: ...Introduction 26 MAC 800 2060026 001C ...
Page 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Page 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Page 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Page 122: ...System Configuration 122 MAC 800 2060026 001C ...
Page 128: ...Maintenance 128 MAC 800 2060026 001C ...
Page 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Page 153: ......