Field
Comment
PDF Naming
On
/
Off
Automatic Naming
Rule.
•
Patient ID
•
Last Name
•
First Name
•
Date of Birth
•
Procedure
•
Date of Test
•
Export Date
Page 4
High Security Mode
Enables/disables high security mode. You
can activate it only if at least one user
with
Edit Users
and
Edit Setup
privileges is
configured with a password.
When
High Security Mode
is enabled, users
are prompted to enter an ID and password
when logging on to the device. You must
add each user in
User Setup
. For more
information, refer to
“User Setup” on page
Audit Trail
Determines whether the device creates an
audit trail of activity. This is available only if
High Security Mode
is enabled and the
CFRA
audit trail option is activated.
For information on activating the CFRA
option, refer to
“Options Setup” on page 115
.
Auto Logoff
Determines whether the device automatically
logs the user off after a predefined period of
inactivity.
See also
Auto Logoff Time
. This is available
only if
High Security Mode
is enabled.
Auto Logoff Time (1255 min)
Determines the length of inactivity, in
minutes, before the device logs off the user.
This is available only if
High Security Mode
is enabled.
Automatically synchronize with Time
Server
Enables/disables automatic synchronization
with an external time server either on the
institution’s network or the Internet. You
must activate a LAN option to set this option.
Time Server Name
Identifies the server with which the device
synchronizes its time. This can be a server on
the institution’s network or on the Internet.
Contact your server administrator for this
information.
Last synchronization at
Display-only field that identifies when the
last synchronization occurred.
Last synchronized from Time Server
Display-only field that identifies where the
last synchronization occurred.
90
MAC™ 800
2060026-001C
Summary of Contents for MAC 800
Page 8: ...8 MAC 800 2060026 001C ...
Page 26: ...Introduction 26 MAC 800 2060026 001C ...
Page 64: ...Recording a Resting ECG 64 MAC 800 2060026 001C ...
Page 70: ...Arrhythmia Mode Recording 70 MAC 800 2060026 001C ...
Page 76: ...RR Analysis 76 MAC 800 2060026 001C ...
Page 122: ...System Configuration 122 MAC 800 2060026 001C ...
Page 128: ...Maintenance 128 MAC 800 2060026 001C ...
Page 138: ...Creating Barcodes 138 MAC 800 2060026 001C ...
Page 153: ......