Revision B
3-17
Maintenance: Domestic Electrical Safety Tests
MAC 5000 resting ECG analysis system
2000657-002
Ground Continuity
This test verifies that there is continuity (less than 100 m
Ω
resistance)
between all the exposed metal surfaces, which have the potential to
become energized, and the ground prong on the mains AC power cord.
If the metal surfaces are anodized or painted, scrape off a small area in
an inconspicuous area for the probe to make contact with the metal.
■
Use a digital multimeter to check ground continuity from the AC
line cord ground pin to exposed metal surfaces. (i.e. rear panel
ground lug, ANA/TTL, and EXT. VID. )
■
If the measurements are significantly out of range, check for
breaks in the power cord or in the internal connections within
the unit.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...