Revision B
4-16
Troubleshooting: Diagnostic Tests
MAC 5000 resting ECG analysis system
2000657-002
Diagnostic Tests
Introduction
Verify that the MAC 5000 resting ECG analysis system operates properly
by running the diagnostic tests. These tests check the operation of the
display screen, speaker, keyboard, thermal writer, battery, and
communication. Detailed information displays on screen.
Loading the System
Diagnostics
1. Select Main Menu on the Resting screen.
2. Select More.
3. Select System Setup.
4. At the prompt type the word “system”, the password set at the
factory, then press the enter key. If the password was not
changed, the System Setup menu appears. If the menu does not
appear, use the master password. If the system’s unique
password is inaccessible, create one following the instructions
in “Substitute Master Password” later in this section.
5. When the System Setup menu displays, hold down shift and
press
F5
(shift +
F5
).
6. Type “prod” at the service password prompt.
7. The System Diagnostics menu appears.
Substitute Master Password
If you do not have access to the system’s password, you can create a
master password as follows.
1. At the prompt for the system password, enter
meimac
. A
random 6-digit number displays on the screen. For example,
876743.
2. Write the number down and create a new 6-digit number by
adding alternating digits from the random number as follows.
Add:
◆
first and third digits,
◆
second and fourth digits,
◆
third and fifth digits,
◆
fourth and sixth digits,
◆
fifth and first digits, and
◆
sixth and second digits.
Disregard the 10s column when adding the digits. The new
number from the example above would be 440020.
3. Enter the new number, then press the enter key. The System
Setup menu displays.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...