Revision B
4-18
Troubleshooting: Diagnostic Tests
MAC 5000 resting ECG analysis system
2000657-002
The second test pattern shows 32 color levels. Check for problems with
the overall pattern. (If the system does not have the color option, various
grey scale patterns display.)
Press the enter key to exit the test.
Anti-Aliasing Test Pattern
This test pattern consists of a large square with a series of lines
projecting from the center of the square to the perimeter of the square.
Specifications not currently defined.
Press the enter key to exit the test. Highlight Return and press the enter
key to return to the System Diagnostics menu.
Speaker Test
Use the arrow pad to select Loud or Soft. Press the enter key to produce a
loud or soft tone. (The tone level difference is minimal.)
Highlight Return and press the enter key to return to the System
Diagnostics menu.
Keyboard Test
Press each key and verify that the key is highlighted on the screen and
also displayed at the top of the screen. (It is normal for a dim
background image to remain on the screen when you select the next
key.) The numeric value that displays at the top of the screen is the scan
code representation of the pressed key.
■
Check both of the shift keys by pressing each in combination
with a letter to display a capital letter.
■
Press the center of arrow pad control and verify that the word IN
displays on screen. Press arrows to change the displayed arrow
position. A beep sounds with each arrow press.
■
Press the shift key and the
F6
key to exit the test.
NOTE
The display shows keys in the upper part of
the screen that are not presently available on
the keyboard.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...