Revision B
4-21
Troubleshooting: Diagnostic Tests
MAC 5000 resting ECG analysis system
2000657-002
■
Battery current
■
Battery temperature
■
Maximum and minimum battery temperature
■
Ambient temperature
■
Maximum and minimum ambient temperature
■
Current battery charging status
Battery Discharge Test
This test charges the battery to full capacity, if necessary, then monitors
a discharge cycle. Monitored information, written to the floppy disk,
includes:
■
Discharge rate (in mAH)
■
Battery temperature
■
Battery charge status
■
Percent of charge remaining
Battery Charge Test
This test completely discharges the battery, if necessary, then monitors
a charge cycle. Monitored information, written to the floppy disk,
includes:
■
Charge rate (in mAH)
■
Battery temperature
■
Battery charge status
■
Percent of charge
Print Discharge Test Results
Print Charge Test Results
Writes the results of the last discharge or charge test to a floppy disk for
later printing to the writer.
Communication Tests
COM Port Loopback Test
The Communications Port Loopback Test sends various ASCII characters
out the COM port’s transmit lines and expects the same character to
return in it’s receive lines.
1. Select COM 1: Base System and press the enter key.
2. Follow the instructions on screen and install loopback jumpers
in the COM 1 port. (Make note of these jumpers as they are the
same for COM 2 Modem Option test.)
While the test is in process, the word Testing appears in the
upper right corner of the screen. Upon completion, the word
Passed or Failed appears, depending on the results.
3. Remove the loopback jumpers when the test is complete.
Follow the same steps for the COM 2 Modem Option test.
Summary of Contents for MAC 5000
Page 1: ...MAC 5000 resting ECG analysis system field service manual PN 2000657 002 Revision A...
Page 2: ...T 2...
Page 3: ...field service manual PN 2000657 002 Revision B MAC 5000 resting ECG analysis system...
Page 12: ...Revision B 1 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 20: ...Revision B 2 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 68: ...Revision B 5 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 88: ...Revision B 6 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 114: ...Revision B 7 2 MAC 5000 ECG analysis system 2000657 002...
Page 123: ...Revision B 7 11 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 125: ...Revision B 7 13 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 127: ...Revision B 7 15 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 129: ...Revision B 7 17 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 131: ...Revision B 7 19 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 133: ...Revision B 7 21 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 135: ...Revision B 7 23 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 137: ...Revision B 7 25 PCB Assemblies SD801212 003A CPU MAC 5000 ECG analysis system 2000657 002...
Page 142: ...Revision B 7 30 PCB Assemblies SD801222 001A LED MAC 5000 ECG analysis system 2000657 002...
Page 146: ...Revision B A 2 MAC 5000 resting ECG analysis system 2000657 002...
Page 152: ...Revision B A 8 MAC 5000 resting ECG analysis system 2000657 002...
Page 154: ...Revision B B 2 MAC 5000 resting ECG analysis system 2000657 002...