Cleaning
Invenia ABUS 2.0 – System Setup and Basic Service Manual
10-7
4700-0043-00 Rev. 4
3. Remove and clean the plastic Clear
Cover of the Scan Head Assembly.
CAUTION:
Be sure to properly seat the plastic Clear Cover securely after cleaning.
4. Use disinfectant wipe to clean all portions of the transducer and Scan Head Assembly that comes in
contact with the patient’s breast. See note on approved disinfectant wipes on page 10-5.
Inspect transducer for wear.
CAUTION:
DO NOT use an alcohol-based solution to clean the transducer.
Table 10-3: Cleaning the Scan Head (Continued)
Steps
Illustrations
Summary of Contents for Invenia ABUS 2.0
Page 5: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 3 4700 0043 00 Rev 4...
Page 6: ...i 4 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 7: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 5 4700 0043 00 Rev 4...
Page 8: ...i 6 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 9: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 7 4700 0043 00 Rev 4...
Page 10: ...i 8 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 11: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 9 4700 0043 00 Rev 4...
Page 12: ...i 10 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 13: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 11 4700 0043 00 Rev 4...
Page 14: ...i 12 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 15: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 13 4700 0043 00 Rev 4...
Page 16: ...i 14 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 26: ...i 24 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 277: ......
Page 278: ...1 2 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...