Diagnostics/Troubleshooting
7-36
Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Hardware
problems
A hardware is an error/malfunction in either the transducer or
the main unit.
A hardware issue may be an error/malfunction on a card.
Software problems
Check if a newer software version is available. A software
update may include noise fixes. If needed, update the software.
The following messages may appear during the scan. Please
refer to the cause/action information provided when trying to
resolve the issue.
Table 7-5: Invenia ABUS 2.0 Messages
System Message
Cause/Action to Take
[Patient Name] is already selected for this scan
You have already selected a patient for this scan.
Abort fininshed. Ready for next scan.
You can start the next scan; the previous scan was
aborted.
Aborted
The scan has been aborted.
Aborting
The current scan is being ended.
Are you sure you want to logout?
Prompts the user to confirm that they intended to
stop using the system.
Dataset is not available
The selected patient’s dataset is no longer available.
It cannot be re-sent to the workstation.
Imaging Engine Initializing....
The system is in the process of starting up.
Please check the scantype
The user must select a scan type before scanning
can begin.
Please input patient information.
Prompts the user to fill in all patient information fields
prior to starting the scan.
Please mark all nipple positions
The nipple positions on the scan must be set before
proceeding with further action on the system.
Please select a new patient
You need to select a new patient in order to begin
the scan.
Please unlock Scan Head Assembly first
Login to begin scanning.
Please wait until processing is finished
The system is still processing the current task;
please wait to start the next tast.
Press compression button
Apply compression before starting the scan.
Press Compression button or Start button
Apply compression, then press Start to begin the
scan.
Press Scan Button on Scan Head to Start Scanning
Press Start on the Scan Head Assembly to begin the
scan.
Press Start button to start scan
Press Start on the Scan Head Assembly to begin the
scan.
Summary of Contents for Invenia ABUS 2.0
Page 5: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 3 4700 0043 00 Rev 4...
Page 6: ...i 4 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 7: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 5 4700 0043 00 Rev 4...
Page 8: ...i 6 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 9: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 7 4700 0043 00 Rev 4...
Page 10: ...i 8 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 11: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 9 4700 0043 00 Rev 4...
Page 12: ...i 10 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 13: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 11 4700 0043 00 Rev 4...
Page 14: ...i 12 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 15: ...Invenia ABUS 2 0 System Setup and Basic Service Manual i 13 4700 0043 00 Rev 4...
Page 16: ...i 14 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 26: ...i 24 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...
Page 277: ......
Page 278: ...1 2 Invenia ABUS 2 0 System Setup and Basic Service Manual 4700 0043 00 Rev 4...