background image

1.3.5  2285053 – Through bolts with insert

Table 2-7:  

Item

Name

Part #

Description

Quan.

Notes

Bolt, Hex

2296892

Through Bolt; M20 - 500-400

12 bolts for Gantry and 4 bolts for

Table

16 pc

Through Bolt mounting

method hardware

Washer, Flat 99142204

Special Flat Washer for Through

Bolts; one for each bolt

16 pc

same as above

Plate

2290941

Special Steel Spacer Plate; 4 in. x

4 in. (102 mm x 102 mm); one for

each bolt

16 pc

same as above

Nut, Hex

99141607

Hex Nut M20 STL galvanized two

for each bolt

32 pc

same as above

Grout assy

2285055

- 10 kg Powdered Mortar Ardex

K-15

- RTV Silicon Rubber Adhesive

- 18 mm making tape adhesive

- Open cell foam

1 kit

Used in constructing Gantry

grout dam

Dowel

2290937

Wood Dowel; 24 mm diameter

12 pc

 

Water Elec‐

tric Separa‐

tor

2268647

Manufactured part introduced in

Positioner key hole to fit 1 electrical

conduit of 6”.

1 pc

Electrical and water con‐

duits separated inch size.

1.3.6  2285632 – Floor plate / Base plate assembly

Table 2-8:  

Item

Name

Part #

Description

Quan.

Notes

Hex Screws 2360523

Screw M16x30x30 Inox A4-80

Pass

12 pc

For floor plate / Base plate

assembly

Washer

99125091

Washer P 17 mm/40 mm

34 pc

24 are used for the position‐

er integration

Hex Screws 99133570

Screw M16x40x40 Inox A4-80

Pass

10 pc

4 only are used for Omega

NOTE:

The kit 2285632 Floor plate / Baseplate assembly is needed when the floor plate is

installed. This part isn’t GE part. It’s ordered locally.

Single Plane and Biplane Innova Systems Pre–Installation Kit Installation Procedures

GE Healthcare

Direction 2290880-12-1EN, Revision 1

44

1  LC Positioner and Table Floor Preparation Kits

Summary of Contents for Innova Series

Page 1: ...GE Healthcare Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures OPERATING DOCUMENTATION 2290880 12 1EN Revision 1...

Page 2: ...Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 2...

Page 3: ...od Ne poku avajte servisirati opremu ako niste u potpunosti pro itali i razumjeli ovaj ser visni priru nik Zanemarite li ovo upozorenje mo e do i do ozljede davatelja usluge operatera ili pacijenta us...

Page 4: ...en alvorens deze onderhoudshandleiding werd geraadpleegd en begrepen is Indien deze waarschuwing niet wordt opgevolgd zou het onderhoudspersoneel de operator of een pati nt gewond kunnen raken als gev...

Page 5: ...es dangers lectriques m caniques ou autres WARNUNG DE Diese Serviceanleitung existiert nur in englischer Sprache Falls ein fremder Kundendienst eine andere Sprache ben tigt ist es Aufgabe des Kunden f...

Page 6: ...olo dopo aver consultato il pre sente manuale ed averne compreso il contenuto Il mancato rispetto della presente avvertenza potrebbe causare lesioni all addetto alla manutenzione all operatore o ai pa...

Page 7: ...wym i zrozumienia go Niezastosowanie si do tego ostrze enia mo e doprowadzi do obra e serwisanta operatora lub pacjenta w wyniku pora enia pr dem elektrycznym zagro enia me chanicznego b d innego ATEN...

Page 8: ...Tento n vod na obsluhu je k dispoz cii len v angli tine Ak z kazn kov poskytovate slu ieb vy aduje in jazyk ako angli tinu poskytnutie prekladate sk ch slu ieb je zodpovednos ou z kazn ka Nepok ajte...

Page 9: ...m jeziku mora stranka za gotoviti prevod Ne posku ajte servisirati opreme e tega priro nika niste v celoti prebrali in razumeli e tega opozorila ne upo tevate se lahko zaradi elektri nega udara mehans...

Page 10: ...Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 This page left intentionally blank 10 Important Information...

Page 11: ...of 2290880 7 1EN The previous release of this document was 2290880 6 100 Updated with modifications to Baseplate kit flow diagram 58 2290880 8 1EN July 2008 Table for table LC positioner updated 58 22...

Page 12: ...Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 This page left intentionally blank 12 Revision History...

Page 13: ...se plate 26 2 4 4 Base plate placement Floor plate 27 3 PIST0004 Above Grade Base Plate Insert Installation 28 4 PIST0005 Floor Anchor Mounting On Grade 31 5 PIST0006 Vascular Above Without Insert 34...

Page 14: ...285054 Templates 45 1 4 1 2285056 A C N Templates 45 1 4 2 2186109 EP Template 45 Chapter 3 SYSTEM FLOOR MOUNTING TEMPLATE 49 1 Templates 49 Single Plane and Biplane Innova Systems Pre Installation Ki...

Page 15: ...a or InnovaIQ Table with the correct alignment to the gantry Isocenter aligned and proper clinical usage distance between gantry and table NOTE The preferred method of gantry table floor mounting is t...

Page 16: ...OF SPECIALLY DESIGNED BENEATH THE FLOOR STEEL CHANNELS TO ENSURE THE THROUGH BOLT MOUNTING HARDWARE DOES NOT FAIL DURING USE THESE CHANNELS MUST BE DESIGNED BY AN AUTHORIZED LOCAL STRUCTURAL ENGINEER...

Page 17: ...acement and floor hole drilling plan The approval should include an inspection of the floor concrete strength identification of any sub floor obstructions and drill hole size and depth requirements NO...

Page 18: ...ble is to be installed This is the isocenter line It passes through the gantry isocenter and the table pivot point Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures...

Page 19: ...o them The template can be slightly adjusted about the cable exit hole isocenter to accommodate the rebar positioning Be sure to shift the isocenter line if necessary to accommodate positioning the Om...

Page 20: ...ench for 3 4 20 mm nuts 1 1 1 4 30 mm socket for 3 4 20 mm bolts 1 15 16 torque wrench socket 1 Vacuum cleaner 1 2 3 2 Consumables Item Qty Effectivity Part Manufacturer Base Plate Assembly 1 2285048...

Page 21: ...pecialist or lead field engineer to verify that the proper full size floor mounting template is used 1 Assure the following tools for grouting 1 2 heavy duty double insulated electric drill 650 rpm or...

Page 22: ...ied to improve bonding and to reduce moisture lost from the grout during drying NOTE If the floor is tiled the tile must be removed under the area to be primed and grouted Remove as much mastic as pos...

Page 23: ...loor 4 Dampen the area to be grouted with water 2 4 1 5 Grout Preparation Procedure Ardex K 15 when mixed with water becomes a liquid compound which seeks its own level and automatically produces a sm...

Page 24: ...Base plate Placement Floor Anchor Installation 2 4 2 1 Gantry Base plate 1 After grout has hardened remove the foam 2 Smooth off the air bubble shells and clean off the bottom of the base plate and b...

Page 25: ...m of the base plate 3 Position base plate over the anchor holes hardstop towards the table Tap each anchor through the base plate and into its hole with a hammer Refer to PIST0002 Prepare Floor LC Bas...

Page 26: ...3 9 6 12 12 holes Refer to Illustration 1 2 5 After you have torqued the bolts go back every half hour and check the torque two times should be adequate A creeping function of the bolt may have caused...

Page 27: ...ert and torque each of the base plate bolts to 20 ft lbs 2 7 daN m to seat the plate and keep it from moving Continue to torque each bolt in 20 ft lbs 2 7 daN m increments until 150 ft lbs 20 daN m fo...

Page 28: ...he insert and the contractor supplied conduit 3 3 Preliminary Requirements 3 3 1 Consumables Item Qty Effectivity Part Manufacturer Base Plate Assembly 1 2285048 grout anchor bolts water electric Sepa...

Page 29: ...Illustration 1 3 Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 Chapter 1 Installation Procedure 29...

Page 30: ...alization No finalization steps Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 30 3 PIST0004 Above Grade Base Pl...

Page 31: ...rbide tipped drill bit for installing gantry and table through bolts 25 mm 1 recommended 1 10 mm 3 8 drill bit for pilot holes 1 Impact drill for use with carbide tipped drill bits We recommend Skill...

Page 32: ...AL STRUCTURAL ENGINEER OR ARCHITECT INSPECT THE COMPRESSION STRENGTH OF THE CONCRETE FLOOR BEFORE USING FLOOR ANCHORS THE MAXIMUM PULLOUT FORCE PER ANCHOR WA CALCULATED ASSUMING A REGULAR WEIGHT CONCR...

Page 33: ...1 5 4 4 Finalization No finalization steps Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 Chapter 1 Installation...

Page 34: ...rill bit for installing gantry and table through bolts 25 mm 1 recommended 1 10 mm 3 8 drill bit for pilot holes 1 Impact drill for use with carbide tipped drill bits We recommend Skill 906 roto hamme...

Page 35: ...SPITAL STRUCTURAL ENGINEER OR ARCHITECT INSPECT THE COMPRESSION STRENGTH OF THE CONCRETE FLOOR BEFORE USING FLOOR ANCHORS THE MAXIMUM PULLOUT FORCE PER ANCHOR WA CALCULATED ASSUMING A REGULAR WEIGHT C...

Page 36: ...5 4 Finalization No finalization steps Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 36 5 PIST0006 Vascular Abo...

Page 37: ...pplied 1 LC Positioner and Table Floor Preparation Kits Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 Chapter 2...

Page 38: ...2 1 Biplane kit selection Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 38 1 LC Positioner and Table Floor Prep...

Page 39: ...cular system floor preparation are contained in catalog There are some additional gantry table mounting kits based on each mounting method through bolts or floor anchors Single Plane and Biplane Innov...

Page 40: ...85059 12 Hole Floor Baseplate 1 pc Lift Plate 2290939 Lifting Straps 2 pc Hex Head Screw 5166535 Hex Head Screw 1 2 13UNC L1 1 2 12 9 Black 2 pc Grease 2295599 Lithium grease lubricant 100g 2 pc Doc 2...

Page 41: ...olt 6 pc Plate 2290941 Special Steel Spacer Plate 4 in x 4 in 102 mm x 102 mm one for each bolt 6 pc Nut Hex 99141607 Hex Nut M20 STL galvanized two for each bolt 12 pc Dowel 2290937 Wood Dowel 24 mm...

Page 42: ...Quan Notes Bolt Hex 2296892 Through Bolt M20 500 400 12 bolts for Gantry and 4 bolts for Table 16 pc Through Bolt mounting method hardware Washer Flat 99142204 Special Flat Washer for Through Bolts on...

Page 43: ...oor anchor with insert Table 2 6 Item Name Part Description Quan Notes Floor Anchor 46 302265P1 Floor Anchor Bolt 5 8 10 x 6 in drilling 12 anchors for Gantry and 4 an chors for Table 16 pc Above Grad...

Page 44: ...owel 24 mm diameter 12 pc Water Elec tric Separa tor 2268647 Manufactured part introduced in Positioner key hole to fit 1 electrical conduit of 6 1 pc Electrical and water con duits separated inch siz...

Page 45: ...ry Omega IV Compact Or Omega V Long Table System Floor Mounting Template Gantry Omega V Long Table System Floor Mounting Template Optional in Chapter 3 Templates for details 1 4 2 2186109 EP Template...

Page 46: ...2359194 Angio Configuration Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 46 1 LC Positioner and Table Floor P...

Page 47: ...ration Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 Chapter 2 Innova LC Positioner and Omega Innova IQ Table F...

Page 48: ...e and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 This page left intentionally blank 48 1 LC Positioner and Table Floor Prepar...

Page 49: ...Floor Mounting Template 1 Templates Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 Chapter 3 System Floor Mount...

Page 50: ...Illustration 3 1 Hole Location in concrete floor Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12 1EN Revision 1 50 1 Templates...

Page 51: ...not necessary to drill the baseplate cover fixing holes H as they are not used Single Plane and Biplane Innova Systems Pre Installation Kit Installation Procedures GE Healthcare Direction 2290880 12...

Page 52: ...Omega V non motorized Long 1278 mm 50 3 in NA 1595 mm 62 8 in InnovaIQ Table 1278 mm 50 3 in 1395 mm 54 9 in 1595 mm 62 8 in NOTE Distance not authorized for new installations Single Plane and Biplan...

Page 53: ......

Page 54: ...2011 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare 283 rue de la Mini re 78530 Buc FRANCE www gehealthcare com...

Reviews: