5.7 Compartment Air Probe Repairs
Refer to “Figure 5-48 Compartment Air Probe Disassembly”.
Compartment air probe
Probe housing cover
Break plate
Probe cable clip
Back panel
O-ring
Humidity
sensor
Cylindrical spacer
Figure 5-48
Compartment Air Probe Disassembly
1. Using a 2.5 mm hex key, remove the screw that secures the probe cable clip to the back wall.
2. Remove the 2 screws (2mm hex key) that secure the probe housing cover and remove it.
3. Disconnect the connectors for the temperature and humidity sensor (if installed) wires and remove the
compartment air probe.
NOTE:
Some early units have compartment air probes with jack connectors at the probe panel and should
be disconnected there. For these units you must order kit number 6600-0241-850 to replace the probe and
wire harness.
4. Feed the probe cable through the hole in the back wall and remove the front of the probe housing.
5. Remove the 4 screws in the break plate and remove the plate.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
6600-0343-000 104
121
Chapter 5: Repair Procedures