B.4 Operating Environment
Temperature
20 to 30ºC
Humidity
10 to 95% Non-condensing relative humidity
B.5 Storage Conditions
Temperature
-25 to 60ºC
Humidity
5 to 95% Non-condensing relative humidity
Pressure
50 to 106 kPa
B.6 User Control Settings
Patient control temperature
35-37.5ºC in 0.1ºC increments
Air control temperature
20-39ºC in 0.1 increments
Radiant heat power
0-100% in 5% increments
Humidity: Servo- % relative humidity
30-95% in 5% increments
B.7 Alarms
High Air Temp
1.5ºC over AST (air set temperature)
Low Air Temp
3.0ºC under AST
Baby Hot
1.0ºC* over BST (baby set temperature)
Baby Cold
1.0ºC* under BST
Fan Failure
Failure of blower system
Air Temp >3ºC
>38ºC for AST <37ºC
Air Temp >40ºC
>40ºC air temperature for AST >37ºC
Air Probe Failure
Compartment air probe failure
Air Probe Disconnect
Disconnection of compartment air probe
Power failure
Power switch on but no power
System failure
Non-recoverable system failure
Add Water
Humidifier water level low
Scale: Weight Exceeds Maximum
Greater than 8kg load (visual only)
Scale: Scale Failure
Detectable system failure (visual only)
* Can be re-set on the Service screen to 0.05ºC
236
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Appendix B: Specifications
Summary of Contents for Giraffe OmniBed
Page 10: ...x 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Table of Contents ...
Page 28: ...xxviii 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved About this Manual ...
Page 48: ...20 6600 0343 000 104 2001 by Datex Ohmeda Inc All rights reserved Chapter 2 Service Checkout ...