5.4.3 Right Rail Internal Repairs
5.4.3.1 Removing the Inner Rail Assembly
Refer to “Figure 5-26 Removing the Lift Motor”, “Figure 5-28 Right Upright Disassembly”, and “Figure 5-44
Heater Door Cable Adjustment”.
Pinch guard
Right upright
Lift rail
Belt channel
Buoyancy springs
Rollers
Side roller bracket
Motor bracket
Access hole
Belt block
Figure 5-28
Right Upright Disassembly
1. Using a 2.5 mm hex key, remove the 2 screws located at either end of the upright in the rear dovetails
that secure belt channel to the inside of the upright. Using a 3mm hex key, remove the side roller bracket.
Remove the lift motor coupler components if still attached to the rail assembly. Pull the entire rail assembly
up out of the upright.
2. Remove the pinch guard by sliding it up then pulling it off.
96
6600-0343-000 104
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.
Chapter 5: Repair Procedures