Table 4: Warmer indicator status
Status
Warmer
covers
Disposable
cartridge
Power LED
indicator
Temperature
LED indicator
Audible
indicator
Description
Action required
Ready
Open or closed
None
Flashing Green
every 3 seconds
Unlit
None
Warmer unit has power, but
is not heating
None
Device power up
Closed
In place
Red
½ second
duration
Red
½ second
duration
½ second beep
Successful device power
up and over temperature
circuit test
Observe. If the LED does
not flash red, replace the
Warmer; and contact
Technical Support.
In operation
Closed
In place
Solid Green
Flashing Blue
None
Infusate temperature is
< 33 °C
None
In operation
Closed
In place
Solid Green
Solid Blue
None
Infusate temperature is
≥ 33 °C and < 35 °C
None
In operation
Closed
In place
Solid Green
Solid or Flashing
Blue >30 seconds
None
Warmer is unable to
heat the infusate within
operational range. Infusate
temperature is
< 20 °C and/or the flow
rate is > 200 mL/min.
Reduce the flow rate if
possible. If there is no
change in operational
temperature, consider
replacing the Warmer and
contact Technical Support.
In operation
Closed
In place
Solid Green
Solid Green
None
Infusate temperature is
≥ 35 °C and ≤ 42 °C
None
In operation
Closed
In place
Solid Green
Solid Yellow
None
Infusate (and/or ambient
temperature) is > 42 °C
but less than an Over Temp
condition
Observe. This state whereby
the infusate is > 42 ºC
should only be entered
periodically during changes
in flow rate or infusate
temperature.
In operation
Closed
In place
Flashing Red
N/A-Varies
Continuous
audible bursts
Internal failure in the
Warmer
Replace the Warmer if this
occurs, and contact Technical
Support.
Continuous
operation
Closed
In place
Solid Green
Flashing Red
High priority
alarm
Continuous
audible bursts
Infusate (and/or ambient
temperature) is > 45 °C
signifying an Over Temp
condition
Stop the fluid flow, and slide
the Warmer covers open to
stop warming. Replace the
Warmer if this occurs, and
contact Technical Support.
Refer to Appendix C for the chart on “Warming system response by temperature.”
Refer to Warnings for additional information.
Cleaning the enFlow IV fluid/blood warming system components
Caution
• Do not clean with:
Ketones (MEK, acetone, etc.)
Abrasive cleaners
• Do not sterilize the Warmer with:
Steam sterilization (autoclave)
Dry heat
• Do not disinfect or sterilize the Controller.
• Do not spray or pour cleaning solutions directly on the Controller.
• Do not allow cleaning solutions to accumulate on the Controller.