- 36 of 40 -
4400-0024-EN Rev A
enFlow
W arm er
T herm om eter
T em perature
probe
inserted into
extension set.
W aste water
container
IV Line set
Extension set
enFlow
Controller
Power
source
Pump
100
mL/min
T em perature
probe
inserted into
extension set.
Extension set
.
5 L
W ater source
50 °C
3. Change the source fluid temperature to 50 °C ± 2 °C. Take extra care while working with hot fluids.
4. Power on the enFlow system.
5. Allow at least 20 seconds for the power-on self-test to complete and the temperature display to ready a stable temperature. Confirm all LEDs are
illuminated green.
6. Release the clamp and establish a fluid flow of 100 ± 20 mL/min.
7. With the thermometer, confirm the temperature of the fluid at both the input and output end of cartridge.
8. The Over-Temperature alarm sounds within approximately 20 seconds or less of the output fluid temperature reaching that of the input fluid level.
A. Confirm the Controller produces an audible beep and displays a red Over-Temp message on the display (Controller only).
B. Confirm the temperature LED on the Warmer flashes red indicating that the fluid has gone over-temperature.