4400-0024-EN Rev A
Purpose
The enCheck Tester was developed to quickly and reliably trigger the over-temperature alarm condition on the enFlow Warmer. Within seconds, the enCheck
unit will heat the Warmer to an over- temperature scenario causing the alarm to sound. Additionally, the enCheck is designed to verify the Warmer operation
at enFlow’s installation site.
Procedure
When the enCheck is connected and running in the normal mode, the heat is generated from the Warmer unit using the same technology as when a
cartridge is installed. This mode allows for confirmation of the temperature output of the Warmer.
1. Plug the Controller into a hospital grade outlet.
2. Connect the enCheck to the Controller by inserting the male plug end of the enCheck Hubbell connector into the female receptacle on the front face of
the Controller. Push it in so that the plug cover is tight against the receptacle.
enFlo
3. Next, connect the Warmer to the enCheck by inserting the male plug end of the Warmer into the enCheck female receptacle.
4. Insert the temperature probe connector on the enCheck into a thermometer. Set thermometer to “K” type setting.