4400-0024-EN Rev A
There are three LEDs on the enCheck.
1. Lighted Orange LED indicates the power is on.
2. Lighted Green LED indicates the enCheck is in the Normal Mode.
3. Lighted Red LED indicates the enCheck is in the Overheat mode.
There is a slide switch for switching the overheating on/off.
There is a thermocouple adapter for connecting to a calibrated thermometer for verification.
Environmental Requirements
Performance
Operating Temperature Range: -5 ˚C to 40 ˚C
Input Voltage: 12-30 V--- 0.5A
Storage Temperature Range: -30 ˚C to 70 ˚C
Maximum of 14 W
Relative Humidity Range: 10% to 90%
Cleaning the enCheck
1. Use only approved cleaning solutions. (Please reference “Cleaning the enFlow IV fluid/blood warming system components” for a list of approved cleaning
solutions.)
2. Moisten a clean cloth with the cleaning solution; do not spray or pour cleaning solutions directly onto the enCheck.
3. Wipe the surface of the enCheck, taking care not to leave excess residual cleaner on the enCheck. If fluid ingress is detected, set the enCheck aside for an
extended period of time to allow it to dry.