- 23 of 40 -
4400-0024-EN Rev A
Guidance and Manufacturer’s Declaration – Emissions
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment specified below. The customer or user of the enFlow 100 with enFlow 121 should
ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
Radiated
CISPR 11
Group 1, Class A
The enFlow 100 with enFlow 121 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Harmonics
IEC 61000-3-2
Class A
The enFlow 100 with enFlow 121 is suitable for use in all establishments, other than domestic, and those
directly connected to the public low voltage power supply network that supplies buildings used for domestic
purposes.
Flicker IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The enFlow 100 with enFlow 121 is intended for use in the electromagnetic environment specified below. The customer or user of the enFlow 100 with enFlow 121 should
ensure that it is used in such an environment.
Immunity Test
EN/ IEC 60601
Test Level
Compliance Level
Electromagnetic Environment – Guidance
ESD
EN/IEC 61000-4-2
Level 3 (±6kV)
Contact
Level 3 (±8kV)
Air
Level 3 (±6kV)
Contact
Level 3 (±8kV)
Air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the
r/h should be at least 30%.
EFT
EN/IEC 61000-4-4
±2kV Mains
±1kV I/Os
±2kV Mains
±1kV I/Os
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
EN/IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage
Dips/Dropout
EN/IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
100% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
See Note 1
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the enFlow 100 with enFlow 121 requires
continued operation during power mains interruptions, it is recommended
that the enFlow 100 with enFlow 121 be powered from an uninterruptible
power supply or battery.
Power Frequency
50/60Hz
Magnetic Field
EN/IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be that of a typical commercial
or hospital environment.
Note: During the 5 second event, the enFlow 100 and 121 power off, but return to normal operation as soon as power is restored.