[EN] English -
5
K395 Issue 3
Example 2) Set up values for
x:y Scaling
= 0 to 100%.
Table 5:
(Part of table)
Menu options - Settings (Output)
Table 5:
(Part of table) Menu options - Settings (Output)
Table 6:
Additional data for Settings (Output):
Settings
x:y
1
Settings
x:y
2
Edit
3
4
▲
▼
■
■
▲
▼
■
■
Sk1 = Edit
= Flow scaling
(mA, pressure only)
Options
Hz
Description
... Units
Frequency Units
. To select one of these units:
Hz: Range
<
1000Hz
kHz: Range 1 ... 50kHz
Pulses/minute (cpm)
Pulses/hour (cph)
Amplitude
To set the amplitude and the mode of the output
signal. Amplitude = 5V (Default).
Bipolar mode
[
✓
]/[ ]: Set this option to make the
signal pass thru zero.
= Unipolar
=
Bipolar
Waveform
To set the waveform for the output signal:
,
,
= Sine, square, or triangle
To select and set up a value for the “Nudge” output.
Example: 0.010 kHz increments.
Additional data: Select Settings
(
■
■
)
To select and set up values for the “Span Check”
output. Example output cycle:
This cycle repeats
automatically.
Additional data (
Table 6
): Select Settings
(
■
■
)
% Step
To select and set up values for the “% Step” output.
Example output cycle:
Auto Repeat - Optional
Additional data (
Table 6
): Select Settings
(
■
■
)
Options
Hz
Description
... Step
To select and set up values for the “Defined Step”
output. Example output cycle:
Auto Repeat - Optional
Additional data (
Table 6
): Select Settings
(
■
■
)
To select and set up values for the “Ramp” output.
Example output cycle:
Auto Repeat - Optional
Additional data (
Table 6
): Select Settings
(
■
■
)
Options
Pulses
Description
Counts
: To set the total number of output pulses.
Rate
: To set the frequency of the output.
Amplitude
: To set the amplitude of the output
signal. Amplitude = 5V (Default).
Bipolar mode
[
✓
]/[ ]: Set this option to make the
signal pass thru zero.
= Unipolar
=
Bipolar
Item
Value
Span Check
Low (0%)
Set the 0% value.
High (100%)
Set the 100% value.
Dwell (d)
Set the period (Hours:Minutes:Seconds) between
each change in value.
% Step
Low (0%), High (100%), Dwell (d): As above.
Step Size (s) ...
%
Set the change in value for each step as a
percentage of the full-scale range (High - Low).
Defined Step
Low (0%), High (100%), Dwell (d): As above.
Step Size (s)
Set the change in value for each step as a
frequency value.
Ramp
Low (0%), High (100%), Dwell (d): As above.
Travel (t)
Set the period (Hours:Minutes:Seconds) to go from
the Low (0%) value to the High (100%) value.
Auto Repeat
If applicable, select this item to repeat a cycle
continuously.
Summary of Contents for Druck DPI 841
Page 4: ......
Page 18: ......
Page 32: ......
Page 46: ......
Page 60: ......
Page 74: ......
Page 87: ...Druck DPI 841 842 频率校验仪和 频率回路校验仪 用户手册 K395 EN DE ES FR IT PT CN JP RU ...
Page 88: ......
Page 101: ...Druck DPI 841 842 周波数校正器 周波数ループ校正器 ユーザーマニ ュ アル K395 EN DE ES FR IT PT CN JP RU ...
Page 102: ......
Page 114: ...12 JP 日本語 K395 第 3 刷 ...
Page 116: ......
Page 130: ......