- [EN] English
K395 Issue 3
To start - Set up the settings
When the task is set up (Table 2/3), use the
Settings
menu
to adjust the input and/or output operation.
If there is additional data for a menu option, select
Settings
(
■
■
) to see the values that are set up. If
necessary, adjust the values. Refer to “Edit functions”.
Table 4:
(Part of table) Menu options - Settings (Input)
Table 4:
(Part of table) Menu options - Settings (Input)
To start - Edit functions
Example 1) Set up a label for
x:y Scaling
= %.
Display: Task
Hz + mA
1
Settings selection
(If applicable)
2
3
■
■
▲
▼
Sk1 = Start/Stop
Sk2 = Settings
Menu:
Settings
4A
5A
6A
Settings
x:y
▲
▼
■
■
(Table 4)
[
✓
]/[ ]
Sk1 = Edit
Menu:
Settings
4B
5B
▲
▼
(Table 5)
[
●
]/[ ]
Options
(If applicable)
Description
... Units
Frequency Units
(Hz input only). To select one of
these units:
Hz: Range
<
1000Hz
kHz: Range 1 ... 50kHz
Pulses/minute (cpm)
Pulses/hour (cph)
... Units
UPM only = “Pressure Units” if you select an IDOS
task (Table 2/3). Select one of the fixed units of
measurement (psi, mbar ... ).
Trigger level
To set the amplitude at which the instrument
senses a frequency signal. Default = 5V.
Auto Detect
[
✓
]/[ ]: Set this option to make the
instrument calculate the value from the available
signal.
To include maximum, minimum and average
values for the measurement task.
Options
(If applicable)
Description
To select and set up a tare value for the
measurement task (a specified value or the reading
on the display).
The instrument subtracts a positive tare value, and
adds a negative tare value.
Additional data: Select Settings
(
■
■
)
To select and set up a scale of values: One local
scale for each measurement task (Maximum: 5).
Additional data (Example 1/2): Select Settings
(
■
■
)
To select and set up the filter values to give a
smoother output for the measurement task:
Band as a % of full scale (FS). The filter
compares each new value with the previous
value. If the new value is outside the band, it is not
filtered.
Low pass filter time constant in seconds.
Increase the value to increase damping
factor.
Additional data: Select Settings
(
■
■
)
To select and set up the alarm values for the
measurement task (maximum and minimum).
Additional data: Select Settings
(
■
■
)
UPM only.
Gage sensors or sensors with differential
operation. A zero correction that makes the
instrument read zero at local pressure.
Leak Test only. To set an applicable period for the
leak test (Hours:Minutes:Seconds).
Settings
x:y
1
Edit
2
Edit
3
■
■
▲
▼
■
■
Sk1 = Edit
A3 - item [27]
Edit
...
Edit
4
5
%_
▲
▼
■
■
A3 - item [27]
Shift
OK
Summary of Contents for Druck DPI 841
Page 4: ......
Page 18: ......
Page 32: ......
Page 46: ......
Page 60: ......
Page 74: ......
Page 87: ...Druck DPI 841 842 频率校验仪和 频率回路校验仪 用户手册 K395 EN DE ES FR IT PT CN JP RU ...
Page 88: ......
Page 101: ...Druck DPI 841 842 周波数校正器 周波数ループ校正器 ユーザーマニ ュ アル K395 EN DE ES FR IT PT CN JP RU ...
Page 102: ......
Page 114: ...12 JP 日本語 K395 第 3 刷 ...
Page 116: ......
Page 130: ......